Validity and reliability of an iPhone App to assess time, velocity and leg power during a sit-to-stand functional performance test.
Fecha
2018-01Disciplina/s
Actividad Física y DeporteTerapia y Rehabilitación
Materia/s
AgeLeg strength assessment
Physical performance
Resumen
The purposes of this study were: (i) Analyze the concurrent validity and reliability of an iPhone App for measuring time, velocity and power during a single sit-to-stand (STS) test compared with measurements recorded from a force plate; and (ii) Evaluate the relationship between the iPhone App measures with age and functional performance. Forty-eight healthy individuals (age range: 26-81 years) were recruited. All participants completed a STS test on a force plate with the movement recorded on an iPhone 6 at 240 frames-per-second. Functional ability was also measured using isometric handgrip strength and self-paced walking time tests. Intraclass correlation coefficients (ICC), Pearson's correlation coefficient, Cronbach's alpha (α) and Bland-Altman plots with 95% confidence intervals (CI) were used to test validity and reliability between instruments. The results showed a good agreement between all STS measurement variables; time (ICC=0.864, 95%CI=0.77-0.92; α=0.926), velocity (ICC=0.9...